Home You are here: Home - Products - Power Analyzer - Power Analyzer

AC310
Single-Phase Power Analyzer

The AC310 Single-Phase Power Analyzer is a high-performance and reliable instrument. It offers voltage and current accuracy of 0.1%, with maximum voltage of 1000 Vrms and maximum current of 20 A, and provides measurement bandwidth from 0.5 Hz to 100 kHz.

Applications:

Standby power consumption and power analysis for single-phase household and commercial appliances, Power analysis for power‑electronics equipment, transformers and generators, Switching power supply power analysis, Power analysis for lighting and LED products

Model

Model Voltage Current Accuracy Interface Size Inquiry
AC310 1000Vrms 20A ±0.1% RS-232/RS-485/Digital I/O 2U
No models match the selected filters

*Specifications for the product series are subject to change without notice.

Product Features

•High accuracy: 16-bit high-speed high-precision AD conversion, basic accuracy up to 0.1%, fastest display update period of 100 ms

• Wide power measurement range: measurable current up to 50 A, minimum power resolution of 0.1 mW, supporting both standby power consumption and rated power measurement

• Wide bandwidth: supports both AC and DC signals; power measurement bandwidth DC, 0.5 Hz ~ 100 kHz. Suitable for power measurement of loads with standard and non-standard sinusoidal waveforms

• Line filtering: built‑in 500 Hz and 5.5 kHz low‑pass filters for measuring fundamental values of PWM waveforms and suppressing high‑frequency current interference from switching power supplies

• Sensors: ratio setting function, compatible with conventional I-I and V-V voltage/current transformers; supports I-V current sensors with a BNC interface, with maximum input voltage of 10 V. High-current sensors are available as an option

• Energy Accumulation: Supports independent accumulation of forward, reverse, and total energy for billing and net‑metering applications

• Threshold alarm: configurable threshold limits for voltage, current, power, power factor and other parameters;automatic upper/lower‑limit detection with alarm output


Dimensions

Model

AC310

Product Name

AC310

Measurement Parameters

Voltage (U), Current (I), Active Power (P), Reactive Power (Q), Apparent Power (S), Power Factor (PF), Voltage Frequency (fU), Current Frequency (fI), Phase Angle (Φ), Efficiency (η), Total Energy (Wh), Forward Energy (Wh+), Reverse Energy (Wh–), Current Integration (Ah), 50th Harmonic Analysis (HDF), Voltage and Current Total Harmonic Distortion (THD), Peak Voltage (Upk), Peak Current (Ipk), Voltage Crest Factor (CfU), Current Crest Factor (CfI), DC Voltage (Udc), DC Current (Idc), Average rectified value (Umn/Imn/Urmn/Irmn)

Input Impedance

Voltage: approx. 2 MΩ

Current Direct Input: Approx. 2.5 mΩ (50 A range)

Current Sensor Input: Approx. 100 kΩ

A/D Sampling Rate

Approx. 100 kS/s

Full-Scale Crest Factor

3

Rated Voltage Range

15/30/60/100/150/300/600/1800*[V]

 (Direct Input)

*Crest factor at 1800 V full scale is 1.5

Rated Current Range

50 A model: 500 m / 1 / 2 / 5 / 10 / 20 / 40 / 50* [A]

 (Direct Input)

*Crest factor at the 50 A range full scale is 1.5

Rated Current Range

50m/100m/200m/500m/1/2/5/10[V]

(Sensor input)

Voltage/Current Accuracy Range

(1%~110%)*× range

*For 1800 V voltage range and 50 A current range: (1% ~ 100%) × range

Power Factor Accuracy Range

±(0.001 ~ 1.000)

Voltage Measurement Accuracy

DC

±(0.1%×indicating value+0.1%× range)

0.5Hz≤f<45Hz

±(0.1%× indicating value+0.2%× range)

45Hz≤f≤66Hz

±(0.1%× indicating value+0.1%× range)

66Hz<f≤1kHz

±(0.1%× indicating value+0.2%× range)

1kHz<f≤10kHz

±({0.1+0.05×(f-1)}%× indicating value+0.2%× range)

10kHz<f≤100kHz

±({0.5+0.04×(f-10)}%× indicating value+0.3%× range)

Current Measurement Accuracy

DC

±(0.1%×indicating value+0.1%×range)

0.5Hz≤f<45Hz

±(0.1%×indicating value+0.2%×range)

45Hz≤f≤66Hz

±(0.1%×indicating value+0.1%×range)

66Hz<f≤1kHz

±(0.1%×indicating value+0.2%×range)

1kHz<f≤10kHz

±((0.1×f)%×indicating value+0.2%×range)

10kHz<f≤100kHz

±({1+0.08×(f-10)}%×indicating value+0.3%×range)

Active Power Measurement Accuracy

DC

±(0.1%×indicating value+0.1%×range)

0.5Hz≤f<45Hz

±(0.3%×indicating value+0.2%×range)

45Hz≤f≤66Hz

±(0.1%×indicating value+0.1%×range)

66Hz<f≤1kHz

±(0.2%×indicating value+0.2%×range)

1kHz<f≤10kHz

±({0.2+0.1×(f-1)}%×indicating value+0.2%×range)

10kHz<f≤50kHz

±({0.2+0.1×(f-1)}%×indicating value+0.3%×range)

50kHz<f≤100kHz

±({5.1+0.18×(f-50)}%×indicating value+0.3%×range)

Active Power Measurement Range

11mW~11kW@220V,PF=0.01~1

Low Power Factor

Apparent power accuracy,±(0.2%×indicating value)@PF=0

Power Accuracy Range

Active power accuracy: add 0.05%@PF=0.001~0.1

Active Power Maximum Resolution

0.1mW

Frequency Measurement Range

DC,0.5Hz ~ 100kHz

Frequency Measurement Accuracy

±0.1%×indicating value

Harmonic Measurement

10 Hz ~ 600 Hz, 1st ~ 50th harmonic content, total distortion

Energy Measurement Range

0 ~ 999,999 MWh (Resolution: 1 mWh / 0.01 mAh)

Energy Measurement Accuracy

±0.2%×indicating value

Expanded Uncertainty

Voltage, current, power, frequency, energy ≤ 0.20%

Filter Function

500 Hz, 5.5 kHz voltage-line, current-line and frequency filtering

Transformation Ratio Range

1 ~ 50000

Data Update Cycle

100m / 250m / 500m / 5[s]

Alarm Functions

Voltage, current, power, power factor; upper/lower limits and threshold settings

Control Interfaces

Standard: RS-232; Optional: RS-485, digital I/O

Overall Dimensions(W×H×D)

213mm(W, front panel)×88mm(H, front panel)×386mm(D,  incl. terminals)

Panel Cutout Dimensions

210(W)×85(H) mm

Foot Height

15 mm

Net Weight

Approx.3kg


Product Documentation

1
AC310 Single-Phase Power Analyzer
Chat Inquiry Contact E-mail
TOP

Product Catalog