The AC310 Single-Phase Power Analyzer is a high-performance and reliable instrument. It offers voltage and current accuracy of 0.1%, with maximum voltage of 1000 Vrms and maximum current of 20 A, and provides measurement bandwidth from 0.5 Hz to 100 kHz.
| Model | Voltage | Current | Accuracy | Interface | Size | Inquiry | |
|---|---|---|---|---|---|---|---|
| AC310 | 1000Vrms | 20A | ±0.1% | RS-232/RS-485/Digital I/O | 2U | ||
| No models match the selected filters | |||||||
*Specifications for the product series are subject to change without notice.
•High accuracy: 16-bit high-speed high-precision AD conversion, basic accuracy up to 0.1%, fastest display update period of 100 ms
• Wide power measurement range: measurable current up to 50 A, minimum power resolution of 0.1 mW, supporting both standby power consumption and rated power measurement
• Wide bandwidth: supports both AC and DC signals; power measurement bandwidth DC, 0.5 Hz ~ 100 kHz. Suitable for power measurement of loads with standard and non-standard sinusoidal waveforms
• Line filtering: built‑in 500 Hz and 5.5 kHz low‑pass filters for measuring fundamental values of PWM waveforms and suppressing high‑frequency current interference from switching power supplies
• Sensors: ratio setting function, compatible with conventional I-I and V-V voltage/current transformers; supports I-V current sensors with a BNC interface, with maximum input voltage of 10 V. High-current sensors are available as an option
• Energy Accumulation: Supports independent accumulation of forward, reverse, and total energy for billing and net‑metering applications
• Threshold alarm: configurable threshold limits for voltage, current, power, power factor and other parameters;automatic upper/lower‑limit detection with alarm output
Model | AC310 | |
Product Name | AC310 | |
Measurement Parameters | Voltage (U), Current (I), Active Power (P), Reactive Power (Q), Apparent Power (S), Power Factor (PF), Voltage Frequency (fU), Current Frequency (fI), Phase Angle (Φ), Efficiency (η), Total Energy (Wh), Forward Energy (Wh+), Reverse Energy (Wh–), Current Integration (Ah), 50th Harmonic Analysis (HDF), Voltage and Current Total Harmonic Distortion (THD), Peak Voltage (Upk), Peak Current (Ipk), Voltage Crest Factor (CfU), Current Crest Factor (CfI), DC Voltage (Udc), DC Current (Idc), Average rectified value (Umn/Imn/Urmn/Irmn) | |
Input Impedance | Voltage: approx. 2 MΩ | |
Current Direct Input: Approx. 2.5 mΩ (50 A range) | ||
Current Sensor Input: Approx. 100 kΩ | ||
A/D Sampling Rate | Approx. 100 kS/s | |
Full-Scale Crest Factor | 3 | |
Rated Voltage Range | 15/30/60/100/150/300/600/1800*[V] | |
(Direct Input) | *Crest factor at 1800 V full scale is 1.5 | |
Rated Current Range | 50 A model: 500 m / 1 / 2 / 5 / 10 / 20 / 40 / 50* [A] | |
(Direct Input) | *Crest factor at the 50 A range full scale is 1.5 | |
Rated Current Range | 50m/100m/200m/500m/1/2/5/10[V] | |
(Sensor input) | ||
Voltage/Current Accuracy Range | (1%~110%)*× range | |
*For 1800 V voltage range and 50 A current range: (1% ~ 100%) × range | ||
Power Factor Accuracy Range | ±(0.001 ~ 1.000) | |
Voltage Measurement Accuracy | DC | ±(0.1%×indicating value+0.1%× range) |
0.5Hz≤f<45Hz | ±(0.1%× indicating value+0.2%× range) | |
45Hz≤f≤66Hz | ±(0.1%× indicating value+0.1%× range) | |
66Hz<f≤1kHz | ±(0.1%× indicating value+0.2%× range) | |
1kHz<f≤10kHz | ±({0.1+0.05×(f-1)}%× indicating value+0.2%× range) | |
10kHz<f≤100kHz | ±({0.5+0.04×(f-10)}%× indicating value+0.3%× range) | |
Current Measurement Accuracy | DC | ±(0.1%×indicating value+0.1%×range) |
0.5Hz≤f<45Hz | ±(0.1%×indicating value+0.2%×range) | |
45Hz≤f≤66Hz | ±(0.1%×indicating value+0.1%×range) | |
66Hz<f≤1kHz | ±(0.1%×indicating value+0.2%×range) | |
1kHz<f≤10kHz | ±((0.1×f)%×indicating value+0.2%×range) | |
10kHz<f≤100kHz | ±({1+0.08×(f-10)}%×indicating value+0.3%×range) | |
Active Power Measurement Accuracy | DC | ±(0.1%×indicating value+0.1%×range) |
0.5Hz≤f<45Hz | ±(0.3%×indicating value+0.2%×range) | |
45Hz≤f≤66Hz | ±(0.1%×indicating value+0.1%×range) | |
66Hz<f≤1kHz | ±(0.2%×indicating value+0.2%×range) | |
1kHz<f≤10kHz | ±({0.2+0.1×(f-1)}%×indicating value+0.2%×range) | |
10kHz<f≤50kHz | ±({0.2+0.1×(f-1)}%×indicating value+0.3%×range) | |
50kHz<f≤100kHz | ±({5.1+0.18×(f-50)}%×indicating value+0.3%×range) | |
Active Power Measurement Range | 11mW~11kW@220V,PF=0.01~1 | |
Low Power Factor | Apparent power accuracy,±(0.2%×indicating value)@PF=0 | |
Power Accuracy Range | Active power accuracy: add 0.05%@PF=0.001~0.1 | |
Active Power Maximum Resolution | 0.1mW | |
Frequency Measurement Range | DC,0.5Hz ~ 100kHz | |
Frequency Measurement Accuracy | ±0.1%×indicating value | |
Harmonic Measurement | 10 Hz ~ 600 Hz, 1st ~ 50th harmonic content, total distortion | |
Energy Measurement Range | 0 ~ 999,999 MWh (Resolution: 1 mWh / 0.01 mAh) | |
Energy Measurement Accuracy | ±0.2%×indicating value | |
Expanded Uncertainty | Voltage, current, power, frequency, energy ≤ 0.20% | |
Filter Function | 500 Hz, 5.5 kHz voltage-line, current-line and frequency filtering | |
Transformation Ratio Range | 1 ~ 50000 | |
Data Update Cycle | 100m / 250m / 500m / 5[s] | |
Alarm Functions | Voltage, current, power, power factor; upper/lower limits and threshold settings | |
Control Interfaces | Standard: RS-232; Optional: RS-485, digital I/O | |
Overall Dimensions(W×H×D) | 213mm(W, front panel)×88mm(H, front panel)×386mm(D, incl. terminals) | |
Panel Cutout Dimensions | 210(W)×85(H) mm | |
Foot Height | 15 mm | |
Net Weight | Approx.3kg | |